Learn More
Precision Systems Osmette™ Multi-Osmette Osmometer
Automatic digital osmometers measure osmolality in mOsm/kg H2O. Osmette Multi-Osmette Osmometer uses the freezing point depression method to provide a quick, accurate indication of total concentration. Instruments measure freezing points of solutions to 0.001°C, then directly translate them to measure of concentration.
Supplier: Precision Systems 2430
Description
Easy to Use
- Add sample tube to well, lower operating head to insert wire and thermistor temperature probe into the sample
- Freezing point is detected by the probe, with accuracy of ±0.001°C
- Digital display shows results directly in mOsm/kg H2O
- Entire cycle takes between one-half and one minute per sample
- Fast cooldown, completely zero-stable meter
- Reach operating temperature in <15 min. vs. 40 to 60 min. for other cryoscopic devices
- Eliminate Dewar flasks, ice baths, calculations
- Generate precise results in just 50 to 70 sec.; process 40 to 60 samples per hour
- Minimal training; simple two-step calibration takes less than five minutes
- All models hold calibration, eliminating daily calibration
Multi-Osmette, Model 2430
- 24-position sample turntable for true walk-away operation
- Feature electronic cooling instead of a bath
- Self-calibrating
- Food and beverage industries, chemical manufacturing, pharmaceutical labs, research and quality control
- Clinical applications include diagnosis of renal activity, disease electrolyte imbalance, hyper- and hyponatremia, measuring osmolality of body fluids
Specifications
Digital | |
115 V | |
Each | |
2430 | |
12 in. | |
20 in. | |
51 cm | |
<1000mOsm/kg H2O: ±2mOsm/kg H2O, >1000mOsm/kg H2O: ±0.5%CV at 300mOsm/kg H2O: <1% | |
30 μL | |
18 in. | |
20 x 18 x 12 in. (51 x 46 x 30 cm) |
50/60 Hz | |
Thermoelectric Cooling Block | |
Osmometer | |
Food and beverage industries, chemical manufacturing, pharmaceutical labs, research and quality control, clinical applications include diagnosis of renal activity, disease electrolyte imbalance, hyper- and hyponatremia, measuring osmolality of body fluids | |
30 cm | |
<1 min. | |
Automatic | |
0 to 2000 mOsm/kg H2O | |
46 cm | |
∼3 min. | |
115 V 50/60 Hz |
By clicking Submit, you acknowledge that you may be contacted by Fisher Scientific in regards to the feedback you have provided in this form. We will not share your information for any other purposes. All contact information provided shall also be maintained in accordance with our Privacy Policy.