Labconco™ Precise™ Controlled Atmosphere Glove Boxes
Provide leak-tight environment for contamination-sensitive materials
Supplier: Labconco™ 5220100
Alert:
Desired Oxygen/Moisture Level: The more oxygen-sensitive the work, the lower the oxygen level required. Oxygen levels below 1% (10,000ppm) may be achieved with oxygen and moisture-removal columns. Maintaining levels as low as 1ppm requires a regenerative drying train accessory.
Includes:
- Pair of neoprene gloves (size 9.75, 30 in. [76cm] long, 15mil gauge)
- 22w fluorescent lamp and control switch
- Interior ceiling-mounted electrical duplex receptacle and switch with 5A (115V model)
Requires : Support base (10-358-108, -109; order separately) vacuum source for purging, and inert gas for filling. Contact your Fisher Scientific Representative for assistance.
Description
Ideal for procedures requiring low levels of oxygen (<5ppm) or moisture (<50ppm) such as work with organometallics, alternative energy cells, or hydrophilic chemicals.
Performance
- Meet Class 1 atmosphere containment conditions for oxygen leak rates per ISO 10648-2
- Performance-tested to maintain <5ppm oxygen/<50ppm moisture levels
- One-piece molded, medium-density polyethylene shell
- Dry epoxy-coated steel superstructure, transfer sleeve and interior shelf
- Inner and outer transfer chamber doors are pivoted, counterbalanced and equipped with quick-release latches
- Black, chemical-resistant phenolic core work surface sealed to box floor
- Nominal 13 cu. ft. interior space in main chamber
- 0.25 in. thick removable, laminated safety glass viewing window with neoprene gasket
- Automatic pressure relief valve activates at ±6 in. of water gauge
- Six manual valves with 0.37 in. compression fittings and hose barbs: Main chamber, 4 (2 purge/fill, 2 drying); Transfer chamber: 2 purge/fill
- Main chamber pressure gauge calibrated from -5 to +5 of water gauge
- Transfer pressure chamber gauge from 0 to -.29 in.Hg
- Light and electrical receptacle switches
Order Info
Dimensions Main chamber: 27.5D x 33.5W x 25 in.H (70 x 85 x 64cm); Transfer chamber: 11 nominal I.D. x 12 in.L (28 x 30cm); Window: 26.37W x 12.5 in.H (67 x 32cm); Pressure relief valve: 4 in. nominal dia. (10.2cm); Overall: 31.62D x 52.75W x 40 in.H (80 x 134 x 102cm); Shipping weight: 260 lb. (118kg)
Certifications
ETL listed
Compliance
Meets Class 1 atmosphere containment conditions per ISO 10648-2 test method.
Specifications
Glove Box | |
UL, ETL, CAN/CSA | |
27.5 in. | |
70 cm | |
40 in. | |
102 cm | |
Polyethylene | |
115 V | |
118 kg | |
33.5 in. | |
85.09 cm |
8 A | |
31.6 in. | |
80 cm | |
Procedures requiring low levels of oxygen (<5ppm) or moisture (<50ppm) such as work with organometallics, alternative energy cells or hydrophilic chemicals | |