Promotional price valid on web orders only. Your contract pricing may differ. Interested in signing up for a dedicated account number?
Learn More
Learn More
PureAire Monitoring Systems Dual O2/CO2 Monitor 0 to 25 and 0 to 50,000 ppm
PureAire Monitoring System's Dual O2/CO2 Monitor provides continuous monitoring of oxygen and carbon dioxide. This is ideal for use in confined spaces where low oxygen concentrations could pose a health hazard, from nitrogen gas storage sites to carbon dioxide beverage gas distribution areas.
Supplier: PureAire Monitoring Systems 99143
Description
- No scheduled maintenance zirconium O2 cell and long-life CO2 sensor
- No calibration is required for oxygen
- Self-calibrating CO2 sensor
- Local backlit concentration display
- Built-in audible horn
- Built-in visual LED alarm indicators
- Alarm level 1 for CO2 / Alarm level 2 for O2
- No drift due to environmental or temperature changes
- 3-year warranty
- 24n V DC Power Supply Included
Specifications
Oxygen | |
Optional equipment-PureAire Horn/Strobe, Remote Digital Display, Multi-Channel Controller | |
0 | |
Air Monitoring System | |
-40°F to +140°F | |
Delivers ±0.2% O2 accuracy (± 1% of full range) /CO2 ±300 ppm, ± 3% of measured value | |
Audible 90 dB/Visual-Orange LED alarm indicator for CO2, Red LED alarm indicator for O2 | |
CE | |
3.25 in. | |
Oxygen 0 to 25%, Carbon Dioxide 0 to 50,000 ppm | |
0.75 in. backlit LCD digital display | |
Polycarbonate-General Purpose | |
4.5 in. | |
24 V Power Supply | |
165.1 mm | |
Wall Mount | |
Delivers ± 0.2% O2 accuracy (± 1 % of Full Range)-O2 | |
Analog 4-20 mA output (Active) | |
24 V | |
0.8 kg | |
82.5 mm |
Oxygen 0 to 25%/CO2 0 to 50,000 ppm | |
Oxygen and Carbon Dioxide | |
Diffusion | |
15 sec. (1 sec. average time) | |
-40°C to +60°C | |
Within 1 second of any change in O2 or CO2 | |
250 mA | |
O2 0 to 25%, CO2 0 to 50,000 ppm | |
82.5 mm | |
6.5 in. x 4.5 in. x 3.25 in. | |
24 VDC external power | |
To detect and warn of unsafe oxygen and/or carbon dioxide levels | |
114.3 mm | |
6.5 in. | |
Oxygen 0 to 25%/Carbon Dioxide 0 to 50,000 ppm | |
One user-adjustable Alarm relay for Oxygen and one user-adjustable Alarm relay for CO2. Relays are rated 2 amp 24 VDC / 24 VAC | |
±300 ppm, ± 3% of measured value - CO2 | |
Oxygen -Zirconium Oxide Sensor Cell/Carbon Dioxide -NDIR sensor cell | |
1.6 lb. | |
3.25 in. |